Radiation dosimetry in the fields of health physics and radiation protection is the measurement, calculation and assessment of the ionizing radiation dose absorbed by an object, usually the human body. This applies both internally, due to ingested or inhaled radioactive substances, or externally due to irradiation by sources of radiation.
Internal dosimetry assessment relies on a variety of monitoring, bio-assay or radiation imaging techniques, whilst external dosimetry is based on measurements with a dosimeter, or inferred from measurements made by other radiological protection instruments.
Radiation dosimetry is extensively used for radiation protection; routinely applied to monitor occupational radiation workers, where irradiation is expected, or where radiation is unexpected, such as in the contained aftermath of the Three Mile Island, Chernobyl or Fukushima radiological release incidents. The public dose take-up is measured and calculated from a variety of indicators such as ambient measurements of gamma radiation, radioactive particulate monitoring, and the measurement of levels of radioactive contamination.
Other significant radiation dosimetry areas are medical, where the required treatment absorbed dose and any collateral absorbed dose is monitored, and environmental, such as radon monitoring in buildings.
The International Committee on Radiation Protection (ICRP) guidance states that if a personal dosimeter is worn on a position on the body representative of its exposure, assuming whole-body exposure, the value of Personal Dose Equivalent Hp(10), is sufficient to estimate an effective dose value suitable for radiological protection. Personal Dose Equivalent is a radiation quantity specifically designed to be used for radiation measurements by personal dosimeters. Dosimeters are known as "legal dosimeters" if they have been approved for use in recording personnel dose for regulatory purposes. In cases of non-uniform irradiation such personal dosimeters may not be representative of certain specific areas of the body, where additional dosimeters are used in the area of concern.
A number of electronic devices known as Electronic Personal Dosimeters (EPDs) have come into general use using semiconductor detection and programmable processor technology. These are worn as badges but can give an indication of instantaneous dose rate and an audible and visual alarm if a dose rate or a total integrated dose is exceeded. A good deal of information can be made immediately available to the wearer of the recorded dose and current dose rate via a local display. They can be used as the main stand-alone dosimeter, or as a supplement to other devices. EPD's are particularly useful for real-time monitoring of dose where a high dose rate is expected which will time-limit the wearer's exposure.
In certain circumstances, a dose can be inferred from readings taken by fixed instrumentation in an area in which the person concerned has been working. This would generally only be used if personal dosimetry had not been issued, or a personal dosimeter has been damaged or lost. Such calculations would take a pessimistic view of the likely received dose.
In radiation therapy, three-dimensional dose distributions are often evaluated using a technique known as gel dosimetry.
Medical radiation exposure monitoring is the practice of collecting dose information from radiology equipment and using the data to help identify opportunities to reduce unnecessary dose in medical situations.
Each measure is often simply described as ‘dose’, which can lead to confusion. Non-SI units are still used, particularly in the USA, where dose is often reported in rads and dose equivalent in rems. By definition, 1 Gy = 100 rad and 1 Sv = 100 rem.
The fundamental quantity is the absorbed dose ( D), which is defined as the mean energy imparted by (dE) per unit mass (dm) of material (D = dE/dm) The SI unit of absorbed dose is the gray (Gy) defined as one joule per kilogram. Absorbed dose, as a point measurement, is suitable for describing localised (i.e. partial organ) exposures such as tumour dose in radiotherapy. It may be used to estimate stochastic risk provided the amount and type of tissue involved is stated. Localised diagnostic dose levels are typically in the 0–50 mGy range. At a dose of 1 milligray (mGy) of photon radiation, each cell nucleus is crossed by an average of 1 liberated electron track.
As dose is averaged over the whole organ; equivalent dose is rarely suitable for evaluation of acute radiation effects or tumour dose in radiotherapy. In the case of estimation of stochastic effects, assuming a LNT model, this averaging out should make no difference as the total energy imparted remains the same.
+Radiation weighting factors WR (formerly termed Q factor) used to represent relative biological effectiveness according to ICRP report 103 |
1 |
2.5 + 18.2·e−ln(E)²/6 |
5.0 + 17.0·e−ln(2·E)²/6 |
2.5 + 3.25·e−ln(0.04·E)²/6 |
2 |
20 |
It is difficult to compare the stochastic risk from localised exposures of different parts of the body (e.g. a chest x-ray compared to a CT scan of the head), or to compare exposures of the same body part but with different exposure patterns (e.g. a cardiac CT scan with a cardiac nuclear medicine scan). One way to avoid this problem is to simply average out a localised dose over the whole body. The problem of this approach is that the stochastic risk of cancer induction varies from one tissue to another.
The effective dose E is designed to account for this variation by the application of specific weighting factors for each tissue ( WT). Effective dose provides the equivalent whole body dose that gives the same risk as the localised exposure. It is defined as the sum of equivalent doses to each organ ( HT), each multiplied by its respective tissue weighting factor ( WT).
Weighting factors are calculated by the International Commission for Radiological Protection (ICRP), based on the risk of cancer induction for each organ and adjusted for associated lethality, quality of life and years of life lost. Organs that are remote from the site of irradiation will only receive a small equivalent dose (mainly due to scattering) and therefore contribute little to the effective dose, even if the weighting factor for that organ is high.
Effective dose is used to estimate stochastic risks for a ‘reference’ person, which is an average of the population. It is not suitable for estimating stochastic risk for individual medical exposures, and is not used to assess acute radiation effects.
+Weighting factors for different organs |
0.08 |
0.12 |
0.12 |
0.12 |
0.12 |
0.12 |
0.04 |
0.04 |
0.04 |
0.04 |
0.01 |
0.01 |
0.01 |
0.01 |
0.12 |
National standards laboratories such as the National Physical Laboratory, UK (NPL) provide calibration factors for ionization chambers and other measurement devices to convert from the instrument's readout to absorbed dose. The standards laboratories operates as a primary standard, which is normally calibrated by absolute calorimetry (the warming of substances when they absorb energy). A user sends their secondary standard to the laboratory, where it is exposed to a known amount of radiation (derived from the primary standard) and a factor is issued to convert the instrument's reading to that dose. The user may then use their secondary standard to derive calibration factors for other instruments they use, which then become tertiary standards, or field instruments.
The NPL operates a graphite-calorimeter for absolute photon dosimetry. Graphite is used instead of water as its specific heat capacity is one-sixth that of water and therefore the temperature increase in graphite is 6 times higher than the equivalent in water and measurements are more accurate. Significant problems exist in insulating the graphite from the surrounding environment in order to measure the tiny temperature changes. A lethal dose of radiation to a human is approximately 10–20 Gy. This is 10–20 joules per kilogram. A 1 cm3 piece of graphite weighing 2 grams would therefore absorb around 20–40 mJ. With a specific heat capacity of around 700 J·kg−1·K−1, this equates to a temperature rise of just 20 mK.
Dosimeters in radiotherapy (linear particle accelerator in external beam therapy) are routinely calibrated using ionization chambers or diode technology or gel dosimeters.
Although the United States Nuclear Regulatory Commission permits the use of the units curie, rad, and rem alongside SI units, the European Union European units of measurement directives required that their use for "public health ... purposes" be phased out by 31 December 1985.
Radiation-related quantities